Two automated retinal image analysis systems achieved acceptable sensitivity and false-positive rates for referable retinopathy and are cost-effective alternatives to a purely manual grading approach.
Abstract
Background:
Diabetic retinopathy screening in England involves labour-intensive manual grading of retinal images. Automated retinal image analysis systems (ARIASs) may offer an alternative to manual grading.
Objectives:
To determine the screening performance and cost-effectiveness of ARIASs to replace level 1 human graders or pre-screen with ARIASs in the NHS diabetic eye screening programme (DESP). To examine technical issues associated with implementation.
Design:
Observational retrospective measurement comparison study with a real-time evaluation of technical issues and a decision-analytic model to evaluate cost-effectiveness.
Setting:
A NHS DESP.
Participants:
Consecutive diabetic patients who attended a routine annual NHS DESP visit.
Interventions:
Retinal images were manually graded and processed by three ARIASs: iGradingM (version 1.1; originally Medalytix Group Ltd, Manchester, UK, but purchased by Digital Healthcare, Cambridge, UK, at the initiation of the study, purchased in turn by EMIS Health, Leeds, UK, after conclusion of the study), Retmarker (version 0.8.2, Retmarker Ltd, Coimbra, Portugal) and EyeArt (Eyenuk Inc., Woodland Hills, CA, USA). The final manual grade was used as the reference standard. Arbitration on a subset of discrepancies between manual grading and the use of an ARIAS by a reading centre masked to all grading was used to create a reference standard manual grade modified by arbitration.
Main outcome measures:
Screening performance (sensitivity, specificity, false-positive rate and likelihood ratios) and diagnostic accuracy [95% confidence intervals (CIs)] of ARIASs. A secondary analysis explored the influence of camera type and patients’ ethnicity, age and sex on screening performance. Economic analysis estimated the cost per appropriate screening outcome identified.
Results:
A total of 20,258 patients with 102,856 images were entered into the study. The sensitivity point estimates of the ARIASs were as follows: EyeArt 94.7% (95% CI 94.2% to 95.2%) for any retinopathy, 93.8% (95% CI 92.9% to 94.6%) for referable retinopathy and 99.6% (95% CI 97.0% to 99.9%) for proliferative retinopathy; and Retmarker 73.0% (95% CI 72.0% to 74.0%) for any retinopathy, 85.0% (95% CI 83.6% to 86.2%) for referable retinopathy and 97.9% (95% CI 94.9 to 99.1%) for proliferative retinopathy. iGradingM classified all images as either ‘disease’ or ‘ungradable’, limiting further iGradingM analysis. The sensitivity and false-positive rates for EyeArt were not affected by ethnicity, sex or camera type but sensitivity declined marginally with increasing patient age. The screening performance of Retmarker appeared to vary with patient’s age, ethnicity and camera type. Both EyeArt and Retmarker were cost saving relative to manual grading either as a replacement for level 1 human grading or used prior to level 1 human grading, although the latter was less cost-effective. A threshold analysis testing the highest ARIAS cost per patient before which ARIASs became more expensive per appropriate outcome than human grading, when used to replace level 1 grader, was Retmarker £3.82 and EyeArt £2.71 per patient.
Limitations:
The non-randomised study design limited the health economic analysis but the same retinal images were processed by all ARIASs in this measurement comparison study.
Conclusions:
Retmarker and EyeArt achieved acceptable sensitivity for referable retinopathy and false-positive rates (compared with human graders as reference standard) and appear to be cost-effective alternatives to a purely manual grading approach. Future work is required to develop technical specifications to optimise deployment and address potential governance issues.
Funding:
The National Institute for Health Research (NIHR) Health Technology Assessment programme, a Fight for Sight Grant (Hirsch grant award) and the Department of Health’s NIHR Biomedical Research Centre for Ophthalmology at Moorfields Eye Hospital and the University College London Institute of Ophthalmology.
Contents
- Plain English summary
- Scientific summary
- Chapter 1. Introduction
- Background
- Automated diabetic retinopathy screening
- Development and process of automated retinal image analysis systems
- Image quality assessment
- Image analysis
- Diagnostic accuracy of automated retinal image analysis systems reported to date
- Human grader performance and current screening
- Rationale for the study
- Research aim
- Research objectives
- Chapter 2. Methods
- Chapter 3. Results
- Data extraction from the Homerton diabetic eye screening programme
- Screening performance of EyeArt software
- Screening performance of Retmarker software
- Screening performance of iGradingM software
- Arbitration on subset of screening episodes
- Exploratory analyses of demographic factors on screening performance
- Altered thresholds
- Implementation
- Health economics
- Chapter 4. Health economics: methodology and results
- Chapter 5. Discussion
- Chapter 6. Conclusions and future research
- Acknowledgements
- References
- Appendix 1. Variables exported from OptoMize Data Export Module
- Appendix 2. Reading centre grading form
- Appendix 3. Additional tables referred to in Chapter 3
- List of abbreviations
About the Series
Article history
The research reported in this issue of the journal was funded by the HTA programme as project number 11/21/02. The contractual start date was in December 2012. The draft report began editorial review in November 2015 and was accepted for publication in July 2016. The authors have been wholly responsible for all data collection, analysis and interpretation, and for writing up their work. The HTA editors and publisher have tried to ensure the accuracy of the authors’ report and would like to thank the reviewers for their constructive comments on the draft document. However, they do not accept liability for damages or losses arising from material published in this report.
Declared competing interests of authors
SriniVas Sadda received grants and personal fees from Optos and Carl Zeiss Meditec during the duration of the study.
Last reviewed: November 2015; Accepted: July 2016.
- NLM CatalogRelated NLM Catalog Entries
- Automated Diabetic Retinopathy Image Assessment Software: Diagnostic Accuracy and Cost-Effectiveness Compared with Human Graders.[Ophthalmology. 2017]Automated Diabetic Retinopathy Image Assessment Software: Diagnostic Accuracy and Cost-Effectiveness Compared with Human Graders.Tufail A, Rudisill C, Egan C, Kapetanakis VV, Salas-Vega S, Owen CG, Lee A, Louw V, Anderson J, Liew G, et al. Ophthalmology. 2017 Mar; 124(3):343-351. Epub 2016 Dec 23.
- Diagnostic accuracy of diabetic retinopathy grading by an artificial intelligence-enabled algorithm compared with a human standard for wide-field true-colour confocal scanning and standard digital retinal images.[Br J Ophthalmol. 2021]Diagnostic accuracy of diabetic retinopathy grading by an artificial intelligence-enabled algorithm compared with a human standard for wide-field true-colour confocal scanning and standard digital retinal images.Olvera-Barrios A, Heeren TF, Balaskas K, Chambers R, Bolter L, Egan C, Tufail A, Anderson J. Br J Ophthalmol. 2021 Feb; 105(2):265-270. Epub 2020 May 6.
- A study of whether automated Diabetic Retinopathy Image Assessment could replace manual grading steps in the English National Screening Programme.[J Med Screen. 2015]A study of whether automated Diabetic Retinopathy Image Assessment could replace manual grading steps in the English National Screening Programme.Kapetanakis VV, Rudnicka AR, Liew G, Owen CG, Lee A, Louw V, Bolter L, Anderson J, Egan C, Salas-Vega S, et al. J Med Screen. 2015 Sep; 22(3):112-8. Epub 2015 Mar 5.
- Review Digital Breast Tomosynthesis with Hologic 3D Mammography Selenia Dimensions System for Use in Breast Cancer Screening: A Single Technology Assessment[ 2017]Review Digital Breast Tomosynthesis with Hologic 3D Mammography Selenia Dimensions System for Use in Breast Cancer Screening: A Single Technology AssessmentMovik E, Dalsbø TK, Fagelund BC, Friberg EG, Håheim LL, Skår Å. 2017 Sep 4
- Review The value of digital imaging in diabetic retinopathy.[Health Technol Assess. 2003]Review The value of digital imaging in diabetic retinopathy.Sharp PF, Olson J, Strachan F, Hipwell J, Ludbrook A, O'Donnell M, Wallace S, Goatman K, Grant A, Waugh N, et al. Health Technol Assess. 2003; 7(30):1-119.
- An observational study to assess if automated diabetic retinopathy image assessm...An observational study to assess if automated diabetic retinopathy image assessment software can replace one or more steps of manual imaging grading and to determine their cost-effectiveness
- Enhanced psychological care in cardiac rehabilitation services for patients with...Enhanced psychological care in cardiac rehabilitation services for patients with new-onset depression: the CADENCE feasibility study and pilot RCT
Your browsing activity is empty.
Activity recording is turned off.
See more...