Evidence Table 5: Summary of ambulatory measurement for articles addressing question #1a-c

Study (question)DeviceDaytimeNighttime
TypeNameValidatedDefinitionTime Interval (mins)DefinitionTime Interval (mins)
Helmers, 2000 (c)oscillometricSpaceLabs 90207Y7:00am -11:00pm1511:00pm - 7:00am60
Hoegholm, 1999 (c)oscillometric oscillometricTM-2420, Model 7
TM-2420, Model 6
Y
Y
8:00am - 9:59pm1512:00am - 5:59am30
Inden, 1998 (c)unknownNikon Colin 630N7:00am -11:30pm3011:00pm - 6:30am30
Ironson, 1989 (a)oscillometricSpaceLabs not specifiedunknown9:00am - 11:00pm20can't tell 
Jula, 1999 (a)auscultatoryAccutracker IIN6:00pm - 11:00am1511:00pm - 6:00am30
Khoury, 1992 (a)oscillometricSpaceLabs 90207Y7:00am - 11:00pm1111:00pm - 7:00am60
MacDonald, 1999 (c)oscillometricSpaceLabs 90207Y8:00am - 10:00pm2010:00pm - 8:00pm60
Manning, 1999 (c)unknownMedilog ABPNpatient reported30patient reported30
Martinez, 1999 (c)oscillometricSpaceLabs 90207Y10:00am - 8:00pm1512:00pm - 6:00am15
Martinez, 2001 (c)oscillometricSpaceLabs 90207Y10:00am - 8:00am1512:00am - 6:00am30
Modesti, 1994 (a)oscillometricSpaceLabs 90207Y7:01am - 10:00pm1510:01pm - 7:01am15
Myers, 1995a (c)oscillometric
oscillometric
unknown
SpaceLabs 90202
SpaceLabs 90207
SpaceLabs 5200
Y
Y
unknown
can't tellcan't tellcan't tellcan't tell
Myers, 1995b (a)oscillometric
oscillometric
SpaceLabs 90202
SpaceLabs 90207
Y
Y
can't tell15not measured 
Narkiewicz, 1995 (a)oscillometric
oscillometric
SpaceLabs 90207
TM-2420, Model 7
Y
Y
6:00am - 11:00pm1011:00pm - 6:00am30
Owens, 1999 (c)oscillometricSpaceLabs 90207Y9:00am - 9:00pm309:01pm - 12:59am30
Palatini, 1998 (c)oscillometric
auscultatory
SpaceLabs 90207
TM 2420, Model 7
Y
Y
6:00am - 11:00pm1011:00pm - 6:00am30
Pierdomenico, 1995 (c)oscillometric
oscillometric
SpaceLabs 90202
SpaceLabs 90207
Y
Y
6:00am - 12:00pm1512:00pm - 6:00am30
Sega, 1994 (a)oscillometricSpaceLabs 90207Y7:00am - 11:00pm2011:00pm - 7:00am20
Staessen, 1999 (a)oscillometric
oscillometric
SpaceLabs 90202
SpaceLabs 90207
Y10:00am - 8:00pm3012:00am - 6:00am30
Stergiou, 1998a (a)oscillometricSpaceLabs 90207Ypatient reported20patient reported20
Stergiou, 1998b (b, c)oscillometricSpaceLabs 90207Ypatient reported20patient reported20
Stergiou, 2000 (b, c)oscillometricSpaceLabs 90207Ycan't tell20can't tell20
Thijs, 1996 (a)oscillometric
oscillometric
unknown
SpaceLabs 90202
SpaceLabs 90207
Plus other unspecified
Y
Y
unknown
10:00am - 8:00pm3012:00am - 6:00am30
Tochikubo, 1999 (c)unknownTM-2425unknownpatient reported30patient reported30
Verdecchia, 1992 (c)oscillometric
oscillometric
unknown
SpaceLabs 90202
SpaceLabs 90207
SpaceLabs 5200
Y
Y
unknown
6:00am - 10:00pm15 15
Verdecchia, 1995 (c)oscillometric
oscillometric
unknown
SpaceLabs 90202
SpaceLabs 90207
SpaceLabs 5200
Y
Y
unknown
6:00am - 10:00pm1510:00pm - 6:00am15
Verdecchia, 1996 (c)oscillometric
oscillometric
SpaceLabs 90202
SpaceLabs 90207
Y
Y
6:00am - 10:00pm1510:00pm - 6:00am15
Zachariah, 1988 (a)unknownPressurometer IIIunknowncan't tell7.5can't tell15
Zachariah, 1991 (a)unknownPressurometerunknowncan't tell7.5can't tell15
Zawadzka, 1998 (a, c)auscultatoryTM 2420unknowncan't tell30not measured 

From: Evidence Tables

Cover of Utility of Blood Pressure Monitoring Outside of the Clinic Setting
Utility of Blood Pressure Monitoring Outside of the Clinic Setting.
Evidence Reports/Technology Assessments, No. 63.
Appel LJ, Robinson KA, Guallar E.

NCBI Bookshelf. A service of the National Library of Medicine, National Institutes of Health.